ISO 9001:2000 includes many requirements for control of monitoring
and measuring devices, and two relate directly to the use of calibration
stickers. Neither requires stickers, and our investigations indicate
a global conspiracy is unlikely.
As with many ISO 9001:2000 requirements, folk tales have emerged
in the 20-odd years since the standard was released. The need for
calibration stickers on all devices is one of those folk tales.
To develop a system that makes the best sense for a given organization,
it is important to first understand the requirements, then think
creatively about how to address the requirements.
Calibration Requirements
ISO 9001:2000 section 7.6 states:
- "The organization shall determine...the monitoring
and measuring devices needed to provide evidence of conformity
of product to determined requirements."
- "...measuring
equipment shall...c) be identified to enable the calibration
status to be determined."
Therefore, ISO 9001:2000 requires devices that check product conformity
be identified, and that identification must include a way to determine
if the device has been calibrated.
Not all devices must be, "calibrated or verified." Only
those used to check product conformity must be addressed (the rest
are optional). A decision must be made to determine what devices
those will be, and how everyone will know which devices those are.
After the devices that will be calibrated have been determined,
the methods for identifying the calibration status must be decided.
Slapping calibration stickers on devices is easy and simple enough,
but some devices and environments are not conducive to the use
of stickers. For example, if the device is used in an environment
where the sticker may become unreadable (or fall off), a calibration
sticker may not be a good solution. If the use of sticker will
change the performance of the device (such as a calibrated weight),
a sticker would also not be appropriate.
It is important to keep the precise requirement in mind, which
is, "...measuring equipment shall...c) be identified to enable
the calibration status to be determined." It is up to the
organization to decide what method is best for them, and then demonstrate
the method selected provides adequate control to satisfy the requirement.
Many Valid Methods
A good registrar auditor understands that many different methods
may be used. Some examples from our experience are listed below.
These examples are intended to generate creative thought on methods
other than calibration stickers, and are not presented as approaches
approved by Platinum Registration. All methods must be tested
during registration audits to determine if the controls implemented
and objective evidence presented satisfy the requirements.
Device Availability - The simple availability of the device may
be a method for indicating the device is calibrated. For example,
if the device is not locked and tagged out, it has been calibrated.
Alternately, an organization may make the calibrated version of
smaller devices available, and all uncalibrated versions of the
device are placed beyond use (stored under lock and key, etc.).
Records - Documents stored nearby or easily accessible may provide
the required identification
Calibrate Before Use - This one is easy to say, and harder to
actually prove. Expect a deep investigation and long interviews
with everyone who uses the device. Remember, a record of calibration
results is also a requirement of ISO 9001.
Creative Thinking Foils Conspiracy
While addressing the ISO 9001 requirements, avoid the conventional
wisdom of tradition and folk tales. Have a well worn version
of ISO 9001 handy, and refer to it often with a critical eye
for what the standard specifies. And if you have any questions,
please contact us.
back to top |